• Ensure skin is intact before treatment.
• The device must only be used with the AC adapter provided. If a different AC adapter is used,
harm may occur.
1. Place the controller on a table in front of the user.
2. If necessary, install the blade to the AC adaptor.
If another blade (plug adapter) is required, contact Dermadry’s Customer Service.
3. Plug the AC adapter into the wall and connect it to the controller.
Note: Before connecting the device to the AC adapter, ensure that
it has been at room temperature for at least the last 6 hours.

The controller’s display will light up and start its automatic internal
testing (all the display symbols and green LED will flash once).
The internal testing will take 1-2 seconds to complete, the device will then switch to 'Settings’ mode.
Settings mode
The default treatment profile is set for treatment of the underarms, which has the lowest current strength.
In ‘Settings’ mode, the current setting point (in mA) and the profile will flash on the screen.
Depending on the desired treatment, users can select from the following profiles: underarms, hands or feet.
Each treatment profile has its own settings with specific maximum current level and treatment duration.
Current strength can be modified depending on sweat level.

Active mode
Treatment mode
Place the underarm electrodes in the underarms.

- The green LED will turn on.
- The current will begin to rise slowly towards the setpoint value.
- The treatment timer will begin to decrease. When less than a minute is left, the display will be in seconds.
- The current setpoint display will change to show the real-time tension in volts. L5 will be displayed when the tension is lower than 5 volts.
During the treatment, the device will automatically alternate the current polarity. Refer to Section
13 ‘Integrated Safety Features’ for more details.
The treatment can be paused or stopped at any time as per the instructions of Section 9 – End of Treatment.
Note: If the tap water does not contain enough minerals, the treatment will not be as effective. Use non-carbonated
bottled or mineral water instead. If this does not work, contact Dermadry’s Customer Service. immediately.
Slight pain could be felt at the beginning of the treatment or after the polarity alternating sequence.